Section 5: Commissioning
05/11 38 51209 Revision 2
5.4.8 Determining the Scale Interval
WP A/Calibration
Max 3000 d
3000 kg
Scale interval 3000 d
1kg
The scale interval
1, 2, 5 10, 20 or 50 can be set by pressing
.
The weight unit is taken from [Max] and cannot be changed here.
The number of digits behind the decimal point must be determined already when entering [Max] as
well.
Set Scale interval
After pressing or confirmation of the change is displayed
with:
Set scale interval failed
Max not multiple of scale
interval
The maximum capacity is not an integer multiple of the scale
interval.